Welcome to the official website of Shanghai Shangcai Testing Machine Co., Ltd.!
XHBT-3000Z III Fully Automatic Three-Indentation Digital Display Brinell Hardness Tester

- The structure implements a 1:1 high-precision gravity-sensing system, coupled with zero-backlash ball screws for force control. A 0.03-class force sensor is employed, Ensure that every load is accurate and error-free;
- The instrument is equipped with 10 Class test force, 13 Brinell hardness test scales, suitable for testing a wide range of metals. and nonmetal Materials;
- At the same time, it also possesses 3 a ball press head and 2 pieces The objective lens automatically switches between the selected scale, objective, and indenter; the instrument will… According to the settings Measurement and observation using the indenter and objective lens. , When measuring different specimens, you no longer have to worry about frequently changing the indenter. It will also prevent damage to the indenter caused by a mismatch between the indenter and the test force. , with a tightly sealed optical design and high magnification, it delivers exceptionally clear images of the surface being observed;
- Turret Positioning is switched mechanically. Limit position and Photoelectric limit switch Dual coordination, high positioning accuracy , ensure the turret’s five-point centering ;
- Hardness tester and micro- All-in-one computer, equipped with Win dows The system possesses all the functions of a computer;
- Built-in CCD image automatic measurement system, 12 inches Touchscreen operation, indentation length and Hardness value Data results are saved in real time and displayed in a list; reports can be generated in WORD or EXCEL format. ;

It is suitable for hardness testing of materials such as cast iron, steel, nonferrous metals, and soft alloys, and also for determining the hardness of certain nonmetallic materials, including hard plastics and bakelite.


X HB T -3000 Z III fully automatic The Brinell hardness tester features a precision‑engineered design, with test force application directly controlled by a load cell, resulting in a compact unit and smooth, accurate force application and release. It is equipped with a built-in touchscreen. and micro- The computer makes parameter settings and result displays more intuitive, simplifies operation, and eliminates the irritation and visual fatigue caused by the light source, thereby reducing measurement errors. The instrument is equipped with three indenter heads and a dual-objective lens. Select Test Once the scale is initiated, the durometer will automatically select. Corresponding The indenter and objective lens, along with the stage, automatically rise. After contacting the indenter, it automatically returns to the focal plane; the instrument then automatically applies load, holds the load, unloads, and switches to the objective lens. The indentation is clearly visible. 、 Automatic measurement, realizing Brinell hardness testing. One click Automation.
- The monolithic cast-iron structure achieves 99% body‑shell rigidity, minimizing deformation, and the housing undergoes more than six months of natural aging to ensure stable physical properties.
- The structure implements a 1:1 high-precision gravity-sensing system, coupled with zero-backlash ball screws for force control. A 0.03-class force sensor is employed, Ensure that every load is accurate and error-free;
- After heat treatment, the lead screw achieves a hardness of 35–40 HRC, which ensures stable measurement data across different lifting heights. Thanks to the combined effects of heat treatment and wire‑drawing processes, the lead screw is robust and highly wear‑resistant, significantly reducing the likelihood of clearance caused by long‑term wear that could compromise measurement stability.
- The instrument is equipped with 10 Class test force, 13 Brinell hardness test scales, suitable for testing a wide range of metals. and nonmetal Materials;
- At the same time, it also possesses 3 a ball press head and 2 pieces The objective lens automatically switches between the selected scale, objective, and indenter; the instrument will… According to the settings Measurement and observation using the indenter and objective lens. , When measuring different specimens, you no longer have to worry about frequently changing the indenter. It will also prevent damage to the indenter caused by a mismatch between the indenter and the test force. , with a tightly sealed optical design and high magnification, it delivers exceptionally clear images of the surface being observed;
- Turret Positioning is switched mechanically. Limit position and Photoelectric limit switch Dual coordination, high positioning accuracy , ensure the turret’s five-point centering ;
- The automatic lifting platform features a precision-designed structure and excellent stability, With jog operation on the worktable, it can be quickly raised or lowered. ;
- Hardness tester and micro- All-in-one computer, equipped with Win dows The system possesses all the functions of a computer;
- Built-in CCD image automatic measurement system, 12 inches Touchscreen operation, indentation length and Hardness value Data results are saved in real time and displayed in a list; reports can be generated in WORD or EXCEL format. ;
- The instrument is equipped with USB port, VGA Interfaces and network ports allow for external connections to networks and other devices, providing users with a wider range of optional features. 。

- Basic Functions : Includes all functions of the image processing and measurement system, such as image acquisition, calibration, image processing, geometric measurement, document annotation, and album management. and Fixed-dose printing Wait ;
- Automatic measurement : Automatically capture the indentation, measure its diameter, and calculate the corresponding hardness value. , supports manual fine-tuning ;
- Platform Lifting: The platform can be configured for fast or slow movement, with jog control and one‑button automatic operation.
- Hardness Value Conversion : The system can display the measured Brinell hardness value. HB W Convert to other hardness values such as HV, HR, etc.;
- Data Statistics: The system automatically calculates statistical metrics such as the average and variance of the measured hardness values.
- Super Poor Alarm: Automatically flags abnormal values and triggers an alarm when the hardness exceeds the specified threshold.
- Test Report : Auto-generated Report in WORD or EXCEL format ;
- Ease of use: With a single click on the interface button, all tasks are completed automatically; manual measurement or result modification is also available if needed. , Simply drag the mouse;
- Strong noise resistance: Advanced, reliable image‑recognition technology can accurately detect indentations on complex sample surfaces, and two automatic measurement modes are available to handle extreme conditions.
- Automatic calibration : The system is equipped with a calibration function, facilitating the measurement of indentation size and allowing for calibration at any time. When paired with a calibration grid, it can be used to ensure accurate measurements. The system can automatically perform full‑field calibration at the intersection points of the calibration grid, thereby eliminating measurement errors caused by lens distortion.

|
Model number |
X HB T -3000 Z III |
|
|
Test force |
kgf |
62.5kgf, 100kgf, 125kgf, 187.5kgf, 250kgf, 500kgf, 750kgf, 1000kgf, 1500kgf, 3000kgf |
|
N |
612.9N, 980.7N, 1226N, 1839N, 2452N, 4903N, 7355N, 9807N, 14710N, 29420N |
|
|
Hardness test range |
3.18 ~ 653HBW |
|
|
Method of Applying Test Force |
Automatic (Load/Keep Charge/Unload) |
|
|
Hardness reading |
12-inch computer touch screen, automatically obtains hardness values. |
|
|
Computer Configuration |
CPU: Intel I 5 , Memory: 8 G, Solid-State Drive: 128 G |
|
|
Camera pixel count |
3 million |
|
|
Convert Ruler |
HV, HK, HRA, HRBW, HRC, HRD, HREW, HRFW, HRGW, HRKW, HR15N, HR30N, HR45N, HR15TW, HR30TW, HR45TW, HS, HBS, HBW You can select multiple rulers for simultaneous conversion. |
|
|
Data Output |
Auto-generated Report in WORD or EXCEL format 3 USB ports, 1 VGA port, 1 Ethernet port |
|
|
Objective and indenter switching |
Three objective lenses and dual objectives are recognized by default and switch automatically. |
|
|
Objective lens |
1 × , 2 × |
|
|
Effective field of view |
1 × :6 mm , 2 × :3 mm |
|
|
Minimum unit of measurement |
1 × : 4.6 mu m, 2 × : 2.3 mu m |
|
|
Shelf life |
0~99s |
|
|
Maximum specimen height |
290 mm |
|
|
Distance from the center of the indenter to the machine wall |
150 mm |
|
|
Power supply |
AC220V,50Hz |
|
|
Implementation Standard |
GB/T 231 , ISO 6506, ASTM E10, JIS Z2243 |
|
|
External dimensions |
690 × 310 × 1000 mm |
|
|
Net weight |
200kg |
|

|
Name |
Quantity |
Name |
Quantity |
|
Instrument host |
1 unit |
Attachment Box |
1 piece |
|
Φ200mm flat test bench |
1 piece |
Φ74mm flat test bench |
1 piece |
|
Φ80mm V-shaped test stand |
1 piece |
Standard hardness block 150~250 HBW10/3000 |
1 piece |
|
Standard hardness block 150~250 HBW5/750 |
1 piece |
Power cord |
1 piece |
|
2A fuse |
2 pieces |
3mm hex key |
1 piece |
|
Carrying handle |
4 pieces |
Dust cover |
1 piece |
|
Product Manual |
1 serving |
Certificate of Conformity |
1 serving |
Previous page