Welcome to the official website of Shanghai Shangcai Testing Machine Co., Ltd.!
MHV-10Z/50Z/V3.0 Fully Automatic Vickers Hardness Tester

1. Combines a host unit with a computer for automatic control and automatic data acquisition.
2. High-precision automatic stage, with auto‑focus and the ability to configure various marking paths.
3. Automatic hardness measurement and automatic calculation of carburized layer depth, etc.
4. Automatically generates WORD reports, meeting the needs of various types of test personnel.

Ferrous metals, nonferrous metals, IC Thin sheets, surface coatings, laminated metals; glass, ceramics, agate, gemstones, thin plastics
Materials; hardness testing of the depth and gradient of the carburized layer and the quench-hardened layer.


The MHV‑10Z/50Z/V3.0 fully automatic Vickers hardness tester integrates a range of advanced technologies, including optical imaging, mechanical displacement, electronic control, digital imaging, image analysis, and computer processing. It uses a host computer to control both the Vickers indenter and the automated sample stage, digitizing the hardness indentation and displaying it on the screen. Through automated and manual readings, it accurately measures the microhardness, case depth, coating thickness, and inter‑point spacing of metals, certain non‑metallic materials, and various thin films and coatings. The system can also capture surface topography and produce fixed‑magnification prints. By breaking away from conventional hardness‑testing methods, this system delivers fully automated, high‑precision, and highly repeatable hardness measurements, making it an essential tool for materials analysis.

|
Model number |
MHV-10Z/V3.0 |
MHV-50Z/V3.0 |
|
|
Test force |
kgf |
0.3kgf, 0.5kgf, 1kgf, 2kgf, 2.5kgf, 3kgf, 5kgf, 10kgf |
0.5kgf, 1kgf, 2kgf, 2.5kgf, 3kgf, 5kgf, 10kgf, 20kgf, 30kgf, 50kgf |
|
N |
2.94N, 4.90N, 9.80N, 19.6N, 24.5N, 29.4N, 49N, 98N |
4.90N, 9.80N, 19.6N, 24.5N, 29.4N, 49N, 98N, 196N, 294N, 490N |
|
|
Hardness test range |
1HV~ 3000 HV |
||
|
Method of Applying Test Force |
Automatic (loading) /Protect Load/Unload) |
||
|
Objective and indenter switching |
Automatic switching |
||
|
Convert Ruler |
HK, HRA, HRB W , HRC, HRD, HRE W , HRF W , HRG W , HRK W , HR15N, HR30N, HR45N, HR15T W , HR30T W , HR45T W , HS, HB W |
||
|
Data Output |
Built-in printer Can generate WORD or EXCEL reports and hardness curve charts |
||
|
Hardness reading |
The computer displays the indentation image and performs automatic measurements. |
||
|
Eyepiece |
10 × Digital Display Micrometer Eyepiece |
||
|
Objective lens |
10 × , 40 × |
10 × , 20 × |
|
|
Total magnification |
100 × , 400 × |
100 × , 200 × |
|
|
Eyepiece resolution |
0.25 mu m, 0.06 mu m |
0.25 μm ,0.13 mu m |
|
|
CCD effective field of view |
10 × :680 mu m, 20 × :340 mu m, 40 × :170 mu m |
||
|
Minimum unit of measurement |
10 × :0.1 μm, 20 × :0.0 5 μm, 40 × : 0.025 μm |
||
|
Shelf life |
0~60s |
||
|
Light source |
Halogen light source |
||
|
X-Y Automatic Test Platform |
Size : 11 0×1 10mm; Itinerary : 50 × 50 mm ; Resolution : 0.001m m |
||
|
Maximum specimen height |
165 mm |
||
|
Distance from the center of the indenter to the machine wall |
130 mm |
||
|
Power supply |
AC220V,50Hz |
||
|
Implementation Standard |
ISO 650 7, ASTM E92, JIS Z224 4, GB/T 4340.2 |
||
|
External dimensions |
530×280×630mm , outer carton dimensions 620×450×7 60 mm |
||
|
Machine weight |
Net weight 35k g , gross weight 47kg |
||
Software Features:
System Integration: By means of serial communication with the hardness tester, the system is seamlessly integrated with the device.
Pressure linkage: When the test force of the hardness tester is adjusted, the system detects the change and displays it in real time.
Turret Interlock: The software controls the hardness tester to automatically switch between the objective lens and the indenter, eliminating the need for manual operation.
Loading Control: The software controls the hardness tester to perform loading, eliminating the need for manual operation.
Measurement integration: The software controls the hardness tester turret, applies the load, and directly reads the Vickers hardness value.
Optical linkage: autofocus.
Image Acquisition: Real-time display of hardness images, with the ability to store and print images.
Automatic measurement: Automatically locates the four vertices of the indentation, delivering fast, accurate results, with multiple specialized algorithms to accommodate various indentation shapes. Supports continuous testing, on-the-spot measurement, and pressure application at specified coordinates.
Automatic point detection: Points are placed near the four vertices of the indentation, and the system automatically identifies the optimal vertex, significantly reducing human error.
Diagonal measurement: Select the upper-left and lower-right corners of the indentation to obtain the hardness value.
Four-point measurement: Select the four vertices of the indentation and record the hardness values.
Hardness Conversion: In accordance with national standards, it automatically performs conversions among Brinell, Rockwell, Vickers, and Knoop hardness values, displaying the results in real time.
Graphical and textual report: Automatically records measurement data, generates hardness–depth curves, and allows saving or printing these curves along with all indentation measurement values. It also enables saving or printing indentation images and the current indentation hardness value. All reports are saved as Word documents.
Results Summary: Measurement results from multiple indentations are exported in Excel format, and the number of measurements, maximum hardness, minimum hardness, average hardness, variance, and other statistical metrics are automatically calculated.
Interlocked Control: The system uses serial communication to detect changes in test force, control the hardness tester turret, apply the load, and display the reading directly.
Automatic Displacement: Equipped with a high-precision X‑Y automated stage.
Automatic Identification: Advanced automatic indentation recognition technology reads D1/D2 and HV values within 0.3 seconds.
Stable performance: It can automatically detect indentation marks that are non‑mirror‑polished, unevenly illuminated, or off‑center.
Powerful features: manual readings, automatic readings, hardness conversion, depth‑hardness curves, indentation images, and graphical‑textual reporting.
Ease of use: Calibration is performed using hardness blocks, aligning with users’ everyday habits. A half-day training session is all it takes to get up and running.
Automatic Reading: An innovative automatic reading algorithm enables rapid, highly accurate measurements of various indentations.
Good repeatability: The automated readings exhibit high reproducibility, meeting users’ professional requirements.
Automatic Scanning: Capable of automatically detecting the edges and contours of the sample.
Software specifications:
1. Computer host: Intel i5 processor, 4 GB RAM, 500 GB hard drive, 19-inch LCD monitor
2. Inkjet printer
3. Supported operating systems: Windows XP, Windows 7
4. Digital Imaging System
High resolution: 1.3 megapixels (1280×1024)
High-speed acquisition: 1280×1024 resolution, 25 frames per second;
High definition: Black-and-white images with improved clarity;
Target size: 1/2 inch
5. X-Y Automatic Stage
Tabletop dimensions: 110 × 110 × 41 mm
Maximum travel: 50 × 50 mm
Minimum step size: less than 2 micrometers
Movement speed: Adjustable;
Control modes: manual control, electric control, computer control;
6. Computer-Controlled XY Stage Mode
Positioning and Movement: The stage moves directly to the position specified in the software.
Fixed-point movement: Select any point on the specimen and move it beneath the indenter.
Directional Movement: Click and drag the stage to move it in any of eight directions; the step size is adjustable.
Arbitrary movement: Drag the stage with the mouse in any direction to easily examine the sample surface.
Variable-speed movement: When the platform moves, there are two speed settings—fast and slow—both of which are selectable and adjustable.
Other features: arbitrary origin positioning, automatic reset, mechanical limit switches, and a variety of professional functions to meet diverse requirements.
7. Measurement Method
Automatic Mode — Automatic Stage Movement (X, Y, Z directions) + Automatic Reading
Manual Mode 1 — Automatic Loading + Manual Eyepiece Reticle Measurement
Manual Mode 2 – Manual stage movement + Manual focusing + Automatic/Manual measurement
8. Automatic Reading / Manual Reading
Automatic reading time: approximately 300 milliseconds per individual indentation reading;
Automatic measurement accuracy: 0.1 micrometer;
Automatic measurement repeatability: ±0.8%
Manual readings: manual point selection, automatic point search, four-point measurement, and two-diagonal measurement.
9. Save/Output Results
Store and output measurement data and experimental parameters, including D1, D2, HV, X, Y, etc.;
Store/export the effective case depth profile report;
Store/Output Image;

|
Name |
Quantity |
Name |
Quantity |
|
Instrument host |
1 unit |
10 × Digital Display Micrometer Eyepiece |
1 piece |
|
Vickers diamond indenter |
1 piece |
Objective lens |
2 pieces |
|
Electric Cross-Table Tester |
1 piece |
Power cord |
1 piece |
|
Leveling adjustment screw |
4 pieces |
Level gauge |
1 piece |
|
Fuse 1 A |
2 pieces |
Halogen lamp 12V, 20W |
1 piece |
|
Hex key wrench 2.5mm |
1 piece |
Screwdriver |
1 piece |
|
Standard hardness block |
2 pieces |
Dust cover |
1 piece |
|
Certificate of Conformity |
1 serving |
Product Manual |
1 serving |
|
Joystick |
1 piece |
Desktop computer |
1 unit |
|
CCD camera |
1 set |
Compatible Take over |
1 piece |
|
USB Hardware dongle |
1 piece |
Measurement software |
1 set |
|
Control-specific cable |
1 piece |
RS232 data cable |
1 piece |
|
Electric Platform Control Box |
1 piece |
|
|