Welcome to the official website of Shanghai Shangcai Testing Machine Co., Ltd.!
MHV-1000Z/V3.0 Fully automatic micro-Vickers hardness tester

1. Combines a host unit with a computer for automatic control and automatic data acquisition.
2. High-precision automatic stage, with auto‑focus and the ability to configure various marking paths.
3. Automatic hardness measurement and automatic calculation of carburized layer depth, etc.
4. Automatically generates WORD reports, meeting the needs of various types of test personnel.

Ferrous metals, nonferrous metals, IC Thin sheets, surface coatings, laminated metals; glass, ceramics, agate, gemstones, thin plastics
Materials; hardness testing of the depth and gradient of the carburized layer and the quench-hardened layer.

MHV-1000Z/ V 3.0 The fully automatic micro‑Vickers hardness tester integrates a range of advanced technologies, including optical imaging, mechanical displacement, electronic control, digital imaging, image analysis, and computer processing. It uses a host computer to control both the Vickers indenter and the automated stage, digitizing the hardness indentation and displaying it on the screen. Through automated focusing, automatic readings, and manual measurements, it accurately determines the microhardness of metals and certain non‑metallic materials, as well as the depth of hardened layers, the thickness of coatings, and the distance between two points. The system can also capture surface topography and produce fixed‑magnification prints. By breaking away from conventional hardness‑testing methods, this system delivers fully automated, high‑precision, and highly repeatable hardness measurements, making it an essential tool for materials analysis.

|
Model number |
MHV-1000Z/V3.0 |
|
|
Test force |
girlfriend |
10gf, 25gf, 50gf, 100gf, 200gf, 300gf, 500gf, 1000gf |
|
N |
0.098 N, 0.246 N, 0.49 N, 0.98 N, 1.96 N, 2.94 N, 4.90 N, 9.80 N |
|
|
Hardness test range |
1HV~ 3000 HV |
|
|
Test mode |
HV/HK |
|
|
Method of Applying Test Force |
Automatic (loading) /Protect Load/Unload) |
|
|
Objective and indenter switching |
Automatic switching |
|
|
Convert Ruler |
HK, HRA , HRB W, HRC, HRD , HREW, HRFW, HRG W, HRKW, HR15N, HR30N, HR45N, HR15TW, HR30TW, HR45TW, HS, HBW |
|
|
Data Output |
Built-in printer Can generate WORD or EXCEL reports and hardness curve charts |
|
|
Hardness reading |
The computer displays the indentation image and performs automatic measurements. |
|
|
Objective lens |
10 × (For observation), 40 × (For measurement) |
|
|
Eyepiece |
10 × Digital Display Micrometer Eyepiece |
|
|
Total magnification |
100 × ,400 × |
|
|
Eyepiece resolution |
0.06 mu m |
|
|
CCD effective field of view |
10 × :680 mu m, 40 × :170 mu m |
|
|
Minimum unit of measurement |
10 × : 0.1 μm, 40 × :0.02 5 μm |
|
|
Shelf life |
0~60s |
|
|
Light source |
Halogen light source |
|
|
X-Y Automatic Test Platform |
Size : 11 0×1 10mm; Itinerary : 50 × 50 mm ; Resolution : 0.001m m |
|
|
Maximum specimen height |
90 mm |
|
|
Distance from the center of the indenter to the machine wall |
100 mm |
|
|
Power supply |
AC220V,50Hz |
|
|
Implementation Standard |
ISO 650 7, ASTM E 384, JIS Z224 4, GB/T 4340.2 |
|
|
External dimensions |
480 × 325 × 545 mm , outer carton dimensions 600 ×36 0 × 800 mm |
|
|
Machine weight |
Net weight 31 kilogram , gross weight 44kg |
|
Software Features:
System Integration: By means of serial communication with the hardness tester, the system is seamlessly integrated with the device.
Pressure linkage: When the test force of the hardness tester is adjusted, the system detects the change in test force and displays it in real time.
Turret Interlock: Software-controlled hardness tester performs objective lens switching. — The indenters switch automatically, eliminating the need for manual control of the hardness tester.
Loading linkage: The software controls the hardness tester to perform loading, eliminating the need for manual operation.
Measurement integration: The software controls the hardness tester turret, applies the load, and directly reads the Vickers hardness value.
Optical linkage: autofocus.
Image Acquisition: Real-time display of hardness images, with the ability to store and print images.
Automatic measurement: Automatically locates the four vertices of the indentation, delivering fast, accurate results, with multiple specialized algorithms to accommodate various indentation shapes. Supports continuous testing, on-the-fly measurement, and indentation at specified coordinates.
Automatic point detection: Select points near the four vertices of the indentation, and the system automatically identifies the optimal vertex, significantly reducing human error.
Diagonal measurement: Select the upper-left and lower-right corners of the indentation to obtain the hardness value.
Four-point measurement: Select the four vertices of the indentation and record the hardness values.
Hardness conversion: According to national standards, Brinell hardness is calculated automatically. — Rockwell — Vickers — Conversion among various hardness scales, including the Rockwell scale, with real-time display.
Image-and-Text Report: Automatically records measurement data and generates hardness values. — Depth curve, save or print hardness — Depth curves and all indentation measurement values. Save or print the indentation image and the current indentation hardness value. All reports are organized according to… WORD Save the document.
Result statistics: The measurement results of multiple indentations are presented as… EXCEL Export data in a specified format and automatically calculate metrics such as the number of measurements, maximum hardness, minimum hardness, average hardness, and variance.
Interlocked Control: The system uses serial communication to detect changes in test force, control the hardness tester turret, apply the load, and display the reading directly.
Automatic displacement: configured for high precision X-Y Automatic platform.
Automatic Recognition: Advanced automatic indentation recognition technology, in 0.3 Read out within seconds D1/D2 and High Voltage Value.
Stable performance: It can automatically detect indentations that are non‑mirror‑polished, unevenly illuminated, or off‑center.
Powerful features: manual readings, automatic readings, hardness conversion, depth‑hardness curves, indentation images, and graphical‑textual reporting.
Ease of use: Calibration is performed using hardness blocks, aligning with users’ everyday habits. A half-day training session is all it takes to get up and running.
Automatic Reading: An innovative automatic reading algorithm enables rapid, highly accurate measurement of multiple indentations.
Good repeatability: The automated readings exhibit high reproducibility, meeting users’ professional requirements.
Automatic Scanning: Capable of automatically detecting the edges and contours of the sample.
Software specifications:
1 , Computer host: I5 Processor, Memory 16 G , hard drive 256GB + 1TB, 21 Inch LCD screen
2. Applicable operating system: WIN 7. Windows 10, etc.
3. Digital Imaging System
High resolution: 130 Megapixels ( 1280 × 1024 )
High-speed acquisition: 1280 × 1024 Resolution: 25 frame / seconds;
High definition: Black-and-white images with improved clarity;
Target size: 1/2 inch
4. X-Y Automatic stage
Countertop dimensions: 110 × 110 ×41 millimeter
Maximum travel: 50 ×5 0 millimeter
Minimum step size: 1 Micrometer
Movement speed: Adjustable;
Control methods: manual control, electric control, computer control;
5. X-Y Stage Computer Control Mode
Positioning and Movement: The stage moves directly to the position specified in the software.
Fixed-point movement: Select any point on the specimen and move it beneath the indenter.
Directional Movement: Click and drag the stage with the mouse to move it in any of eight directions; the step size is adjustable.
Arbitrary movement: Drag the stage with the mouse in any direction to easily examine the sample surface.
Variable-speed movement: When the platform moves, there are two speed settings—fast and slow—both of which are selectable and adjustable.
Other features: arbitrary origin positioning, automatic reset, mechanical limit switches, and a variety of professional functions to meet diverse requirements.
6. Measurement Method
Automatic mode —— Automated stage movement ( (X, Y, Z directions) + Automatic reading
Manual mode 1—— Auto-loading + Manual eyepiece reticle measurement
Manual mode 2—— Manually move the stage. + Manual focus + Automatic / Manual measurement
7. Automatic Reading / Manual reading
Automatic reading time: approximately per single indentation reading 300 Milliseconds;
Automatic measurement accuracy: 0.1 micrometer;
Automatic measurement repeatability: ±0.8%
Manual reading: manual point selection, automatic point searching, 4 Point measurement, 2 Diagonal measurement;
8. Save Results / Output
Storage / Output measurement data and experimental parameters, including D1,D2, HV , X , Y Wait;
Storage / Generate a report on the effective case depth profile.
Storage / Output image;

|
Name |
Quantity |
Name |
Quantity |
|
Instrument host |
1 unit |
10 × Digital Display Micrometer Eyepiece |
1 piece |
|
Diamond micro-Vickers indenter |
1 piece |
10 × ,40 × Objective lens |
Each 1 piece |
|
Weight |
6 pieces |
Weight axis |
1 piece |
|
Electric Cross-Table Tester |
1 piece |
Slice holder stage |
1 piece |
|
Flat-Clamp Holder |
1 piece |
Filament holder stage |
1 piece |
|
Leveling screw |
4 pieces |
Level gauge |
1 piece |
|
Fuse 1 A |
2 pieces |
Halogen lamp 12V, 20W |
1 piece |
|
Power cord |
1 piece |
Screwdriver |
2 pieces |
|
Standard hardness block 400~500 HV0.2 |
1 piece |
Standard hardness block 700~800 HV1 |
1 piece |
|
Dust cover |
1 piece |
Certificate of Conformity |
1 serving |
|
Product Manual |
1 serving |
Joystick |
1 piece |
|
Desktop computer |
1 unit |
CCD camera |
1 set |
|
Compatible Take over |
1 piece |
USB Hardware dongle |
1 piece |
|
Measurement software |
1 set |
Control-specific cable |
1 piece |
|
RS232 data cable |
1 piece |
Electric Platform Control Box |
1 piece |
Previous page