Welcome to the official website of Shanghai Shangcai Testing Machine Co., Ltd.!
MHV-10Z/50Z/V2.0 Automatic Vickers Hardness Tester

1. Combines a host unit with a computer for automatic control and automatic data acquisition.
2. High-precision automatic XY stage, capable of configuring various dot‑matrix paths.
3. Automatic hardness measurement, with automatic calculation of carburized layer depth and other parameters.
4. Automatically generates WORD reports, meeting the needs of various types of test personnel.

Ferrous metals, nonferrous metals, IC Thin sheets, surface coatings, laminated metals; glass, ceramics, agate, gemstones, thin plastics
Materials; hardness testing of the depth and gradient of the carburized layer and the quench-hardened layer.

MHV-10Z/50Z/V2.0 The automated Vickers hardness tester integrates a range of advanced technologies, including optical imaging, mechanical displacement, electronic control, digital imaging, image analysis, and computer processing. It uses a host computer to control both the Vickers indenter and the automated stage, digitizing the hardness indentation and displaying it on the screen. Through automatic and manual readings, it accurately measures the microhardness, case depth, coating thickness, and inter‑point spacing of metals, certain non‑metallic materials, and various thin films and coatings. It can also capture surface topography and produce fixed‑magnification prints. This system breaks with conventional hardness‑testing methods, enabling semi‑automatic, high‑precision, and highly repeatable measurements, making it an essential tool for materials analysis.

|
Model number |
MHV-10Z/V2.0 |
MHV-50Z/V2.0 |
|
|
Test force |
kgf |
0.3kgf, 0.5kgf, 1kgf, 2kgf, 2.5kgf, 3kgf, 5kgf, 10kgf |
0.5kgf, 1kgf, 2kgf, 2.5kgf, 3kgf, 5kgf, 10kgf, 20kgf, 30kgf, 50kgf |
|
N |
2.94N, 4.90N, 9.80N, 19.6N, 24.5N, 29.4N, 49N, 98N |
4.90N, 9.80N, 19.6N, 24.5N, 29.4N, 49N, 98N, 196N, 294N, 490N |
|
|
Hardness test range |
1HV~2967HV |
||
|
Method of Applying Test Force |
Automatic (Load/Keep Charge/Unload) |
||
|
Objective and indenter switching |
Automatic switching |
||
|
Data Output |
Built-in printer Can generate WORD or EXCEL reports and hardness curve charts. |
||
|
Hardness reading |
The computer displays the indentation image and performs automatic measurements. |
||
|
Eyepiece |
10 × Digital Display Micrometer Eyepiece |
||
|
Objective lens |
10 × , 40 × |
10 × , 20 × |
|
|
Total magnification |
100 × , 400 × |
100 × , 200 × |
|
|
Eyepiece resolution |
0.25 μm, 0.06 μm |
0.25 μm, 0.13 μm |
|
|
Shelf life |
0~60s |
||
|
Light source |
Halogen light source |
||
|
X-Y Automatic Test Platform |
Dimensions: 110×110 mm; Travel: 50×50 mm; Resolution: 0.002 mm |
||
|
Maximum specimen height |
165mm |
||
|
Distance from the center of the indenter to the machine wall |
130mm |
||
|
Power supply |
AC 220V, 50Hz |
||
|
Implementation Standard |
ISO 6507, ASTM E92, JIS Z2244, GB/T 4340.2 |
||
|
External dimensions |
530×280×630mm, outer carton dimensions 620×450×760mm |
||
|
Machine weight |
Net weight 35 kg, gross weight 47 kg |
||
Software Features:
System Integration: By means of serial communication with the hardness tester, the system is seamlessly integrated with the device.
Pressure linkage: When the test force of the hardness tester is adjusted, the system detects the change in test force and displays it in real time.
Turret Interlock: Software-controlled hardness tester performs objective lens switching. — The indenters switch automatically, eliminating the need for manual control of the hardness tester.
Loading Control: The software controls the hardness tester to perform loading, eliminating the need for manual operation.
Measurement integration: The software controls the hardness tester turret, applies the load, and directly reads the Vickers hardness value.
Optical linkage: manual focusing.
Image Acquisition: Real-time display of hardness images, with the capability to store and print images.
Automatic measurement: Automatically locates the four vertices of the indentation, delivering fast, accurate results, with multiple specialized algorithms to accommodate various indentation shapes. Supports continuous testing, on-the-spot testing, and indentation at specified coordinates.
Automatic point localization: Points are selected near the four vertices of the indentation, and the system automatically identifies the optimal vertex, significantly reducing human error.
Diagonal measurement: Select the upper-left and lower-right corners of the indentation to obtain the hardness value.
Four-point measurement: Select the four vertices of the indentation and record the hardness values.
Hardness conversion: According to national standards, Brinell hardness is calculated automatically. — Rockwell — Vickers — Conversion among various hardness scales, including the Rockwell scale, with real-time display.
Image-and-Text Report: Automatically records measurement data and generates hardness values. — Depth curve, save or print hardness — Depth curves and all indentation measurement values. Save or print the indentation image and the current indentation hardness value. All reports are organized according to… WORD Document saved.
Result statistics: The measurement results of multiple indentations are presented as… EXCEL Export data in a specified format and automatically calculate metrics such as the number of measurements, maximum hardness, minimum hardness, average hardness, and variance.
Interlocked Control: The system uses serial communication to detect changes in test force, control the hardness tester turret, apply the load, and display the reading directly.
Automatic displacement: configured for high precision X-Y Automatic platform.
Automatic Recognition: Advanced automatic indentation recognition technology, in 0.3 Read out within seconds D1/D2 and HV Value.
Stable performance: It can automatically detect indentation marks that are non‑mirror‑polished, unevenly illuminated, or off‑center.
Powerful features: manual readings, automatic readings, hardness conversion, depth‑hardness curves, indentation images, and graphical‑textual reporting.
Ease of use: Calibration is performed using hardness blocks, aligning with users’ everyday habits. A half-day training session is all it takes to get up and running.
Automatic Reading: An innovative automatic reading algorithm enables rapid, highly accurate measurements of various indentations.
Good repeatability: The automated readings exhibit high reproducibility, meeting users’ professional requirements.
Automatic Scanning: Capable of automatically detecting the edges and contours of the sample.
Software specifications:
1 , Computer host: I5 Processor, memory 4G , hard drive 500G,19 Inch LCD screen
2 , inkjet printer
3 , Supported Operating Systems: WIN XP 、WIN7
4 , Digital Imaging System
High resolution: 130 Megapixels ( 1280 × 1024 )
High-speed acquisition: 1280 × 1024 Resolution: 25 frame / seconds;
High definition: Black-and-white images with improved clarity;
Target size: 1/2 inch
5、 X-Y Automated Sample Stage
Countertop dimensions: 110 × 110 × 41 millimeter
Maximum travel: 50 × 50 millimeter
Minimum step size: less than 2 Micrometer
Movement speed: Adjustable;
Control modes: manual control, electric control, computer control;
6、 X-Y Stage Computer Control Mode
Positioning and Movement: The stage moves directly to the position specified in the software.
Fixed-point movement: Select any point on the specimen and move it beneath the indenter.
Directional Movement: Click the mouse to control the stage and move it in any of eight directions; the movement step size is adjustable.
Arbitrary movement: Drag the stage with the mouse in any direction to easily examine the sample surface.
Variable-speed movement: When the platform moves, there are two speed settings—fast and slow—both of which are selectable and adjustable.
Other features: arbitrary origin positioning, automatic reset, mechanical limit switches, and a variety of professional functions to meet diverse requirements.
7 , Measurement method
Automatic mode —— Automated stage movement ( X 、 Y Direction) + Automatic reading
Manual mode 1—— Auto-loading + Manual eyepiece reticle measurement
Manual mode 2—— Manually move the stage. + Manual focus + Automatic / Manual measurement
8 , Automatic reading / Manual reading
Automatic reading time: approximately per single indentation reading 300 Milliseconds;
Automatic measurement accuracy: 0.1 micrometer;
Automatic measurement repeatability: ±0.8%
Manual reading: manual point selection, automatic point searching, 4 Point measurement, 2 Diagonal measurement;
9 , Save Results / Output
Storage / Output measurement data and experimental parameters, including D1,D2, HV , X , Y Wait;
Storage / Generate a report on the effective case depth profile.
Storage / Output image;

|
Name |
Quantity |
Name |
Quantity |
|
Instrument host |
1 Taiwan |
10 × Digital Display Micrometer Eyepiece |
1 only |
|
Vickers diamond indenter |
1 only |
Objective lens |
2 only |
|
Electric Cross-Table Tester |
1 one |
Power cord |
1 root |
|
Leveling adjustment screw |
4 only |
Level gauge |
1 one |
|
Fuse 1A |
2 one |
Halogen lamp 12V、 150,000–200,000 |
1 one |
|
Hex key wrench 2.5mm |
1 to put |
Screwdriver |
1 to put |
|
Standard hardness block |
2 block |
Dust cover |
1 only |
|
Certificate of Conformity |
1 portion |
Product Manual |
1 portion |
|
Printer manual |
1 portion |
Measurement software |
1 set |
|
Desktop computer |
1 Taiwan |
Inkjet printer |
1 Taiwan |
|
CCD Camera |
1 set |
Adaptation takeover |
1 root |
|
USB Hardware dongle |
1 only |
Control-specific cable |
1 root |
|
RS232 Data cable |
1 root |
Electric Platform Control Box |
1 only |
|
Joystick |
1 only |
|
|
Related products